Oncology · Laboratory quality control

Standardised protocols for laboratory quality control

Oncology testing unit · 4 analysers 12 weeks · 4 analysers · 1 shared control

Calibration schedules, proficiency testing and continuous monitoring keep accuracy, precision and reproducibility at a documented standard.

Laboratory work for this case study

The problem

Four analysers produced results that agreed at normal concentrations but drifted apart at the low end of the measuring range, where the clinical decision is actually made. The unit could not show which instrument had moved, because calibration records were kept per operator rather than per analyser.

The method

A shared control material was run on all four instruments each morning for twelve weeks. Results were charted per analyser with Westgard rules applied, calibration events were logged against instrument serial numbers, and proficiency samples were exchanged with a second laboratory to check bias rather than only precision.

The result

Two analysers were shown to drift within eleven days of their calibration interval. Shortening the interval on those units removed the low-end disagreement; the other two kept the original schedule. The unit now holds twelve weeks of documented evidence that its four instruments agree where the result is acted on.